HTM-1103B HRM-45DT Motorized Superficial Rockwell Hardness Tester

Product Detail

Technical Parameter:

1. The Initial Test Force:  29.42N (3kg);  Tolerance: ±2.0%

2. The Total Test Force:  147.1N(15kg),  294.2N(30kg),  441.3N(45kg);  Tolerance: ±1.0%

3. The Allowed Tolerance of Hardness Display Value of the Hardness Tester

Hardness Scale

Hardness Range of the Standard

Hardness Blocks

Max. Tolerance of Hardness Display Value

HR15N

(7094)HR15N

±2HRN

HR30N

(4286)HR30N

HR45N

(2077)HR45N

HR15T

(6793)HR15T

±3HRT

HR30T

(2982)HR30T

HR45T

(1072)HR45T

4. The Indenter Specification:

4.1 The Diamond Rockwell Indenter

4.2 The ф1.5875 mm Ball Indenter

5. The Power Source:     AC220V±5%, 50-60 HZ

6. Time-delayed control:    2-60 seconds, adjustable

7. The Max. Height of the Specimen:  175 mm

8. The Distance from the Indenter Center to the Instrument Body:  165mm.

9. The Overall Dimension of the Tester:     520×215×700mm

10. The Net Weight of the Tester:           73kg

 

Accessories (The Packing List)

No.

Description   of Goods

Quantity

1

Diamond   Rockwell Indenter

1   PC

2

ф1.5875mm Ball Indenter

1   PC

3

Testing   Table (big, medium, V-shaped)

TOTAL   3 PCS

 

Standard   Superficial Rockwell Hardness Block

 

4

HR30T

         1 PC

5

HR15N;   HR30N

TOTAL   2 PCS

6

Fuse   2A

2   PCS

7

Power   Cord

1   PC

8

Weight   A, Weight B, Weight C

TOTAL   3 PCS

9

Level

1   PC

10

Horizontal   Regulating Screw

4   PCS

11

Screwdriver

1   PC

12

Solid   Spanner

1   PC

13

Nylon   Anti-dust Bag

1   PC

14

Usage   Instruction Manual

1   PC

 

Equipment Overview:

1. Hardness is one of the important characteristics which indicate the mechanic performance of material, while the 

hardness testing is an important method to judge the quality of the metal material or its component parts.

2. The Motorized Superficial Rockwell Hardness Tester is adopted a mechanism to automatically load and unload the 

test force, through rotating the Load-Change Hand Wheel to exchange test force, therefore the operation for this 

instrument is very simple, easy and prompt. Besides setting zero to dial, there is no man made error. The instrument 

is with high sensitivity and stability, so as suitable for usage in workshop and testing laboratory.

Show