Technical Parameter:

Model

ETC-2004B Walk-in Constant Temperature and Humidity Testing Chamber (double door)

Inner size(mm)

W1000*H1000*D2000mm( can be specified)

Working environmental temperature

1.Environmental temperature of equipment is 5~35℃, relative humidity ≤85%RH

2.Cycling cooling water temperature of water tower≤30℃, cycle cooling water supply 

0.25~0.4MPa

Temperature range

A、The limit temperature range of machine:-35℃~+115℃

B、The operating temperature range of machine:-30℃~+110℃

Humidity range

40%~98%R.H.(20% humidity corresponds to temperature ranges refer to the attached 

pages for the curve)

Temperature departure

≤±2.0℃

Temperature fluctuation

≤±0.5℃

Relative humidity fluctuation

1.≤ ±3.0%RH(humidity > 75%RH)

2.≤ ±5.0%RH(humidity ≤ 75%RH)

Heating time

1.5℃/ min on average, no linear demand (Under no load situation)

Cooling time 

About 1.0℃/min, no linear demand (Under no load situation)

Interior material

SUS304Mirror stainless steel plate

External material

Assemble cold storage board:colored steel plate with plastic spraying and galvanized,

thermal resistance stress material

Insulation material

High temperature resistance, high density, Chlorine ethyl formate vinegar foam insulation 

material

Weight capacity

Floor of test room can be able to load 200kg/m2 (uniform load)

Cooling system

Water cooling type/ two-stage compressor

Protective device

1. Overtemperature protection

2. Overtemperature protection for humidifying barrel

3. Overload protection for water pump

4. No.1 Compressor protection

5. No.2 Compressor protection

6. Refrigerant high voltage protection

Standard Parts and Accessories

1.Laboratory light:Damp-proof light 80W/220V,4 pieces (two for standard feature, 

two for replacement)

2.Test hole:φ50mm (1) (Can custom the diameter and number of test hole)

3.Wet and dry bulb gauze:5 piece

Controller

“TEMI”Color Touch Screen made in Korea

Door of test room

1.Double point hinges door, with lock function

2.Glass observation window, with electric heating anti-condensation 

(visible range more than 500mm*400mm)

3.Door frame with anti-condensation device

4.Equipped with key for door, can lock the door during high temperature test, 

avoiding accident to hurt staff.

Control panel

1.Temperature(humidity)control display

2.Control the lamp

3.Overtemperature protection device on side

4.Emergency stop switch, running light, fault indicating light,alarm

Heater 

1.Nickel chromium alloy heating wire type heating tube

2.Heater control mode:Non-contact cycle pulse width adjustment, SSR( solid state replay)

Humidifier

1.The humiditying method of humidifier

2.The control method of heater:Non-contact cycle pulse width adjustment, 

SSR( solid state replay)

3.Water level control device, prevent dry heating device for heater

4.Physical principles for humidifying electric heating tube:Apply electric heating tube 

heating to produce steam, then the steam into the climate chamber humidifying the air.

Compressor

 “Tecumseh “ made in Germany

Refrigeration principle

Steam compress type refrigeration

Its principle is to use freon in the latent heat of evaporation of a liquid from cooling objects, 

such as heat and refrigeration. The expansion valve into the evaporator liquid, absorbing 

latent heat of evaporation and evaporation from the surrounding objects, evaporate into 

low temperature and low pressure gas refrigeration compressor suction, compressed by 

compressor into a high temperature and high pressure gas into the water cooled or air 

cooled condenser, which cools into a high-pressure liquid, then through the expansion 

valve into the evaporator. Repeat cycle, 

heat from the low temperature refrigeration and heat released from the high temperature.

Cooling type

Water cooling type:water recycle cooling 15m3/h

Control type of cooling machine

1.PLC for control system (Programmable),according to the testing condition, automatically 

adjust the refrigerator to run.

2.Condensing pressure regulating valve(cooling adjusting device)

3.Evaporating pressure regulating valve (to prevent the evaporator freezing)

4.compressor air cooling circuit

5.Energy loop

Refrigerating fluid

R404A(Ozone depletion index is 0 )


Equipment Description:

Walk-in constant temperature and humidity test room is specially suitable for whole machine and large parts to do low temperature 

test, high temperature test, high and low temperature change test, constant temperature and humidity test, high temperature and

low temperature test with humidity change test and so on.To provide a large temperature and humidity test room for testing big 

size parts,semi-finished product and finished product,at the same time,also be applied in testing large quantity or big size equipment, 

such as computer terminal, auto parts,etc.The test room is mainly consist of control panel, distributor, moisturizing base plate blower, 

heater, humidifier, refrigerating unit.We also can custom the size chamber size and function, according to customers’ standard and 

requirement.Block type appearance,beautiful shape,scientific design,can meet customers’ different requirement. 

 

Following Below Standards:

1.GB/T2423.1-2008 Electric and electronic product environment testing 2nd part: test methods-test a low temperature 

2.GB/T2423.2-2008 Electric and electronic product environment testing 2nd part: test methods-test b: high temperature 

3.GB/T2423.3-2006 Testing 2nd part for electric and electronic products test methods-test Cab:damp heat, steady state testing

4.GB/T2423.4-2008 Electric and electronic product environment testing 2nd part: test methods-test Db alternating hot and humid 

(12h+12h cycle 

5.GB10589-2008 Specification for low temperature test chambers 

6.GB11158-2008 Specification for high temperature test chambers 

7.GB10586-2006 Specification for damp heat test chambers 

8.GJB150.3-1986 High temperature test

9.GJB150.4-1986 Low temperature test

10.GJB150.9-1986 Damp heat temperature test

(Heat capacity of load is not more than 20kg/m3 per cubic metre steel, hot and humid test without under heat load)

Show